On this page
Measurement time and transmission time
Keep transmitter, probe and receiver channel names together. A request to compare temperature or humidity readings can be quoted separately from a check for gaps or timestamp differences in the transmitted data.
If the logger retains records locally during a radio outage and later uploads them, the acquisition time and upload time may differ. For an enquiry about a gap in a report, include both time fields where available and identify whether the concern is the sensing channel or the data transfer.
Specifications and service checks
A missing stretch on a dashboard can result from a communications gap even when the logger has local memory. Compare the on-device export with the receiver history for the same timestamps and note the logging interval. This distinguishes a sensor-reading question from a data-transfer question.
- Range and configuration
- A wireless logger’s radio range describes communication. Its temperature or relative-humidity span belongs to the probe actually connected. Record both probe and logger identities.
- Manufacturers and models
- Vaisala’s RFL100 accepts different probe types. Testo’s 174 H records temperature and humidity locally.
- What we check
- Compare the measured channel with a suitable environmental reference and read back the stored result. Check the remote display uses the same units and channel assignment.
- Reference equipment and accuracy
- Use a calibrated thermometer or humidity reference with uncertainty established for the requested conditions. A second display of the same sensor is not an independent reference.
- When to repeat the check
- Review the interval when a replaceable probe changes. A radio outage calls for a review of missing records and is distinct from a probe calibration issue.
Technical references Vaisala RFL100 guide · Testo 174 data logger comparison