Identify the wetted sensor
Give the sensor bore, connection size and liquid in service. Include the transmitter model, its configured measuring range and the direction the plant calls positive flow. If the transmitter has recently been exchanged, tell us whether the configuration came from the original unit.
What a signal check measures
A comparison with liquid flowing through the measuring tube includes the sensor response. Electrical simulation at the transmitter checks the electronics for the simulated input. These jobs answer different maintenance questions. Name the one required on the order and include the normal low and high operating flow.
Record the result where it is used
A production operator might log cubic metres per hour on the local display while utilities adds up pulses for the daily consumption figure. Which reading goes into your records? If it is the total, include the configured volume per pulse. If it is the analogue signal, include its range and units. For an installed comparison, identify a point on the line that normally runs full.
Electromagnetic flow meter
Two identical-looking meter bodies can have different sensor factors. If a transmitter head has been exchanged, record the factor now entered and the serial numbers on the head and body. That identifies the combination behind the current flow reading.
Choose the boundary of the check. Transmitter simulation tests how the electronics interpret an input. A wet-flow comparison passes liquid through the sensor and includes the measuring tube and electrodes. If the operator relies on a PLC total, identify that receiving channel too. The three checks cover different parts of the line.
| Requested comparison | Parts included |
|---|---|
| Transmitter simulation | Applies a compatible simulated input to check the electronics and output scaling. |
| Wet-flow comparison | Passes liquid through the measuring section and compares its indicated flow with a reference. |
| Receiving channel | Compares the signal accepted by the control system with its displayed units. Identify where the test signal is applied. |
- Range and configuration
- Quote the bore and normal flow in m³/h or L/min. A single flow rate represents a different velocity in a 25 mm line and a 100 mm line.
- Manufacturers and models
- The Endress+Hauser Promag W 400 is a magnetic flowmeter for conductive liquids. Pipe size determines which flow rates correspond to the specified velocity range.
- What we check
- Compare measured volume with a calibrated collection system or reference flowmeter. The tube must be full for a meaningful magnetic-meter comparison.
- Reference equipment and accuracy
- The master can be a calibrated volume prover, weighing system with density correction or suitable transfer meter. Its uncertainty must cover the working rates.
- When to repeat the check
- Review any change in zero or totals after electrode cleaning, relining or process-water conductivity changes.
Technical references Proline Promag W 400 electromagnetic flowmeter · Endress+Hauser electromagnetic measuring principle