Electromagnetic flow meter calibration

On a magnetic flow meter, the sensor in the pipe and the transmitter on the wall may carry different serial numbers. A plant asset register often lists only the transmitter. RTS needs both identities when the order covers the assembled meter.

Send an enquiry
Electromagnetic flow meter — equipment illustration
Equipment illustration
On this page

Identify the wetted sensor

Give the sensor bore, connection size and liquid in service. Include the transmitter model, its configured measuring range and the direction the plant calls positive flow. If the transmitter has recently been exchanged, tell us whether the configuration came from the original unit.

What a signal check measures

A comparison with liquid flowing through the measuring tube includes the sensor response. Electrical simulation at the transmitter checks the electronics for the simulated input. These jobs answer different maintenance questions. Name the one required on the order and include the normal low and high operating flow.

Record the result where it is used

A production operator might log cubic metres per hour on the local display while utilities adds up pulses for the daily consumption figure. Which reading goes into your records? If it is the total, include the configured volume per pulse. If it is the analogue signal, include its range and units. For an installed comparison, identify a point on the line that normally runs full.

Electromagnetic flow meter

Two identical-looking meter bodies can have different sensor factors. If a transmitter head has been exchanged, record the factor now entered and the serial numbers on the head and body. That identifies the combination behind the current flow reading.

Choose the boundary of the check. Transmitter simulation tests how the electronics interpret an input. A wet-flow comparison passes liquid through the sensor and includes the measuring tube and electrodes. If the operator relies on a PLC total, identify that receiving channel too. The three checks cover different parts of the line.

How much of the installed flow measurement needs checking?
Requested comparisonParts included
Transmitter simulationApplies a compatible simulated input to check the electronics and output scaling.
Wet-flow comparisonPasses liquid through the measuring section and compares its indicated flow with a reference.
Receiving channelCompares the signal accepted by the control system with its displayed units. Identify where the test signal is applied.
Range and configuration
Quote the bore and normal flow in m³/h or L/min. A single flow rate represents a different velocity in a 25 mm line and a 100 mm line.
Manufacturers and models
The Endress+Hauser Promag W 400 is a magnetic flowmeter for conductive liquids. Pipe size determines which flow rates correspond to the specified velocity range.
What we check
Compare measured volume with a calibrated collection system or reference flowmeter. The tube must be full for a meaningful magnetic-meter comparison.
Reference equipment and accuracy
The master can be a calibrated volume prover, weighing system with density correction or suitable transfer meter. Its uncertainty must cover the working rates.
When to repeat the check
Review any change in zero or totals after electrode cleaning, relining or process-water conductivity changes.

Technical references Proline Promag W 400 electromagnetic flowmeter · Endress+Hauser electromagnetic measuring principle