Identify the function and polarity
Look at the setting used for the measurement. A current reading is expressed in amperes, an integrated charge in coulombs. List the working points with their sign and units. If your procedure reports the reading after a fixed interval, include that interval as well. A high-resistance or voltage function used on the same bench can be listed separately.
Cables, guards and settling time
Record whether the instrument is used with its own triaxial lead, a fixture or a probe station. A shield and a driven guard serve different purposes, and two cables that fit the socket may be wired differently. Include the connection diagram if the fixture is part of the job. Leakage and settling after a range change matter most at the lowest readings.
Comparing two sets of readings
An input offset recorded before applying a signal helps interpret a later shift. Keep it alongside the range, integration setting and applied value. If you need to compare a result before and after adjustment, request both sets of readings for the function that caused the concern.
Electrometer
Record the cable and guarding configuration used for small currents. If the bench is set up for charge rather than current, state the units and observation interval for that channel.
A very small current reading can change when the input cable is changed. Identify the measuring range, lead and screening arrangement that were in use. For an initial review, a record of the zero indication under the instrument’s specified zero-check condition is particularly informative.
- Range and configuration
- Keithley cites resistance measurements up to 10¹⁷ Ω on its 6517B under its specified method. Current, charge, voltage and derived resistance are separate functions on an electrometer.
- Manufacturers and models
- The Keithley 6517B is an electrometer and high-resistance meter documented by Tektronix.
- What we check
- For the function in use, compare the indication with a suitable electrical source or reference artefact. Identify the range, guarded connection and settling assumptions in the report.
- Reference equipment and accuracy
- Select traceable low-current sources, high-value resistors or charge references as appropriate. Insulation leakage and environmental conditions can dominate uncertainty.
- When to repeat the check
- Review measurements after connector contamination or input repair. Use a stable check artefact near the normal measurement range between calibrations.
Technical references Keithley 6517B user manual