+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Rotameter calibration

A rotameter scale belongs with a particular tube, float and fluid. A reading of five litres per minute on a water scale and the same number on an air scale describe different jobs. The first detail RTS needs is the marking beside the scale.

Send an enquiry
Rotameter — equipment illustration
Equipment illustration
On this page

Read the tube as well as the tag

Some panels have several interchangeable-looking tubes. Record the scale unit, maximum graduation and fluid marked on the tube or plate. Note the float shape and any reference mark used to read its position. A photograph at eye level helps resolve a faded unit or a scale fitted to the panel beside the glass.

Pressure and temperature on a gas scale

Gas flow scales may state the conditions to which their graduations refer. Include these markings and the normal operating pressure when you ask for a quotation. For liquid service, identify the actual liquid and its working temperature. Let us know if the meter has a control valve, alarm switch or separate electrical output.

Working points on a short scale

A process using just two graduations can name both values on the order. If the operator reads a marked band, give the lower and upper marks. Mention a sticking float or an unreadable part of the scale so the team can discuss the condition before comparison.

Rotameter

Read the float at the edge specified for its shape. If two tubes have been exchanged on the same panel, check the tube and float markings as well as the scale printed beside them.

A float with a marked reading edge must be read at the feature specified for that float shape. If two operators disagree by one division, note which part of the float each used and whether the tube was vertical. A tube photo is particularly helpful for identifying the scale.

Range and configuration
KROHNE lists liquid ranges spanning 10 L/h to 120 m³/h across H250 M40 versions. Each float and scale covers a smaller band for a specified fluid and operating conditions.
Manufacturers and models
KROHNE H250 M40 is a metal-tube variable-area flowmeter. Its float and scale are selected as a pair.
What we check
Read the float or pointer at steady reference flows with the meter in its specified orientation. Include points near the normal operating position.
Reference equipment and accuracy
Use a calibrated liquid or gas flow reference and suitable pressure and temperature instruments for any scale correction.
When to repeat the check
Recheck after float replacement or deposits that restrict free movement.

Technical references Technical data and operating conditions of the H250 M40 variable area flowmeter

Enquire about Rotameter

Service enquiry

Service enquiry