On this page
Pressure at the point of use
Copy the unit and pressure reference from the dial or display. If the instrument monitors a vacuum oven, include the pressure at which the oven is released for use. A result at that working point is easier to use than a check near the upper end alone.
A packing line might specify its release limit in mbar absolute while an older record shows a negative value in bar gauge. To compare those figures you need the atmospheric pressure used as the reference as well as the unit conversion.
Specifications and service checks
A vacuum-packaging recipe may call for 150 mbar absolute while the gauge reads in kPa absolute. Those marks represent the same pressure, 15 kPa. Putting both units on the job sheet lets the operator recognise the recipe point when the dial comes back.
- Range and configuration
- WIKA lists full scales down to 25 mbar absolute within the 532.52 family. A gauge open to the room reads atmospheric pressure on an absolute scale.
- Manufacturers and models
- WIKA 532.52 uses an absolute pressure reference. Record the model suffix and absolute pressure unit.
- What we check
- Compare readings against known absolute pressures, including the lower part of the working scale. Record any correction used for the reference level.
- Reference equipment and accuracy
- Use a calibrated absolute pressure standard. A gauge-pressure comparison needs a separate barometric measurement and an uncertainty allowance for it.
- When to repeat the check
- Review the interval against low-end stability and the consequences of a shift in the working vacuum.
Technical references Absolute pressure gauge in stainless steel - WIKA