+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Accelerometer calibration

Accelerometer sensitivity tells you how much electrical output to expect for a given acceleration. A comparison at one frequency gives a sensitivity value there. If the vibration you measure covers a band, readings across that band show how the response changes with frequency.

Send an enquiry
Accelerometer — equipment illustration
Equipment illustration
On this page

Mounting the sensor on the measured axis

Identify the axis, mount, cable and conditioning arrangement used in service. Give the vibration frequencies of interest and whether readings are expressed as peak or RMS. If the result covers an instrument chain, list the readout as well as the pickup.

One vibration channel may be scaled in mV/g while another uses mV/(m/s²). For the same sensor, 98.1 mV/g is approximately 10 mV/(m/s²). Copy the unit as well as the sensitivity from the previous record before changing the value entered in a recorder.

Accelerometers and signal conditioning

Give the sensitivity units, excitation arrangement and frequencies at which results are needed. Include the mounting method from the application because it belongs to the measurement arrangement. RTS measures the accelerometer’s mechanical sensitivity with its mounting and conditioner identified; conditioner-only checks are reported separately.

Specifications and service checks

An accelerometer fitted to a gearbox may be used to measure one narrow vibration band. Identify the shaft frequency or frequency band of interest as well as the sensitivity marked on the sensor. That lets the comparison target the band the condition-monitoring team uses.

Range and configuration
Sensitivity is specified at a reference frequency and changes with frequency. List the acceleration amplitude, frequency band and units actually needed.
Manufacturers and models
PCB Piezotronics 352C33 is a single-axis accelerometer with nominal 100 mV/g sensitivity. Supply the sensor and its conditioning details.
What we check
Compare sensor output and reference acceleration on a vibration exciter over agreed frequencies. Record the mounting arrangement and conditioning used.
Reference equipment and accuracy
Use a traceable reference accelerometer and calibrated acquisition chain. Alignment, fixture response and reference sensitivity contribute to uncertainty.
When to repeat the check
Review after severe shock, a mounting change or suspected cable damage. Retain periodic sensitivity checks at the same frequency.

Technical references PCB Piezotronics 352C33 data sheet

Enquire about Accelerometer

Service enquiry