Air flow calibrator calibration

A flow calibrator used to check a sampling pump has its own flow indication to verify. Identify the gas, flow direction and whether the reading represents actual volume or volume referred to stated temperature and pressure conditions.

Send an enquiry
Air flow calibrator — equipment illustration
Equipment illustration
On this page

Flow heads and sampling rates

A pump used at two different sampling rates may need the low-range head for one and the standard head for the other. List the rate, head serial number and units for each setting. If the pump is programmed in litres per minute and the calibrator reports standard litres per minute, include the reference conditions before asking the lab to compare the figures.

Air flow calibrators

Give its principle, inlet and outlet arrangement, working flow points and the reference conditions applied to the result. Keep pressure and temperature channels separately identified. RTS calibrates the air flow reference as a complete measurement arrangement, including the relevant pressure and temperature conditions.

Specifications and service checks

A sampler is normally set to 2 L/min, although its flow calibrator also covers much larger flows. Identify 2 L/min as the point that matters and copy the volume basis from the sampler. A reading expressed at standard conditions needs the same basis on the comparison sheet.

Range and configuration
Mesa’s Defender low, medium and high flow cells cover 5–500, 50–5,000 and 300–30,000 cm³/min respectively. State which cell is installed and whether the result is actual or normalised flow.
Manufacturers and models
Mesa Labs lists the Defender 510, 520 and 530+ with different measurement and reporting functions.
What we check
Compare the indicated flow with an independent traceable gas-flow standard. Agree gas, direction and reference conditions before comparing corrected values.
Reference equipment and accuracy
Select a transfer standard or primary volumetric reference with uncertainty stated for the gas and flow. Temperature and pressure measurements matter when a volume is normalised.
When to repeat the check
Keep check results for the flow cell itself. Review the interval after a leak, contamination or a change to the pressure or temperature sensor.

Technical references Mesa Labs Defender series · Mesa Defender flow-cell specifications