Actual and standard flow units
Copy the displayed unit in full, including any prefix for normal or standard volume. Give the line pressure, temperature, bore and direction of flow. If the reading is used to allocate air consumption, include the totaliser unit as well as the instantaneous rate.
A compressor house might total air in normal cubic metres while a local instrument shows cubic metres per hour at line conditions. Before comparing the two, identify the pressure and temperature basis and whether one value is accumulated and the other is a rate.
Specifications and service checks
A circular duct with an area of 0.20 m² and a representative mean air speed of 3 m/s carries 0.60 m³/s. A single reading near a duct wall may differ from the cross-sectional mean. Clarify which value the instrument displays before comparing volume-flow figures.
- Range and configuration
- Record whether the scale shows actual litres per minute or litres referred to specified standard conditions. The same mass of air can occupy different volumes at the inlet and outlet.
- Manufacturers and models
- TSI 5200-series meters are designed for low gas flow, including work below 1 L/min.
- What we check
- Compare the indicated flow with a reference at steady points across the operating band, including the smallest flow that matters to the process.
- Reference equipment and accuracy
- A calibrated gas flow standard, temperature measurement and pressure measurement support the comparison. Check the reference specification at the low end.
- When to repeat the check
- Review the interval after a filter blockage or oily air reaches the sensing element.
Technical references TSI 5000-series flowmeter specifications