+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Analytical weight calibration

Two small weights in the same case may have the same nominal mass. They can be distinguished by a dot, a bend in the wire or their marked positions in the case. Keep those details with each piece, including any sheet weights, when sending the set.

Send an enquiry
Analytical weight — equipment illustration
Equipment illustration
On this page

How the laboratory applies weight corrections

List the nominal masses and the class marked on the certificate or case. If the laboratory applies corrections to nominal values, ask for the assigned value required by its weighing procedure. Note any piece replaced since the previous certificate was issued.

A set may contain two 200 mg pieces distinguished by a small mark on one of them. If one has been exchanged with a neighbouring set, keep it listed separately until its individual identity and assigned mass have been reconciled with the register.

Specifications and service checks

Identical nominal masses can have different individual corrections. For a balance check built up from three pieces, record the markings on the actual three so their assigned values can be added in the same combination.

Range and configuration
For milligram pieces, state the individual denomination and identify duplicates in the box. A 10 mg piece cannot be selected as a working standard from its nominal mass alone.
Manufacturers and models
METTLER TOLEDO supplies OIML-class weights. For a box or cast-iron piece, the denomination, class and individual ID matter more than the case model.
What we check
Compare each weight with calibrated reference masses on a suitable comparator. Report its conventional mass or correction and the uncertainty against its individual ID.
Reference equipment and accuracy
OIML R 111 classes run from E1 through F and M classes. Choose references and a comparator that can resolve the tolerance of the weight being tested.
When to repeat the check
Shorten the review interval for loose weights exposed to abrasion, corrosion or cleaning. Set a recall for each replaced or repaired piece.

Technical references METTLER TOLEDO OIML and ASTM Weight Classes: Tolerances, Accuracy, and How to Choose · OIML R 111-1 weight classes

Enquire about Analytical weight

Service enquiry