On this page
Defining the waveform at the output
Name the waveform or provide its defining parameters, the selected output load, amplitude convention and DC offset. For a two-channel sequence, identify the channel timing or phase relationship needed as well as the amplitude on each channel.
A pulse train saved as a fixed number of samples can change duration if the sample clock setting changes. Preserve both the file or waveform identifier and the playback rate. The same named waveform does not on its own establish the interval between pulses.
Specifications and service checks
Two files with identical shapes may be scaled to different peak amplitudes or DC offsets by the output settings. Save the waveform identifier and the amplitude and offset settings with an enquiry about a clipped or shifted production stimulus.
- Range and configuration
- Amplitude can be stated for a matched 50 Ω load or a high-impedance termination. The indicated and observed voltages can differ by a factor of two if those settings disagree.
- Manufacturers and models
- Keysight 33500B-series generators include 20 MHz and 30 MHz models. Identify the output channel and waveform mode.
- What we check
- Verify the sample rate, amplitude and timing of defined features in a saved waveform, together with sine-wave amplitude and frequency where required. Retain the file or pattern identity with the result.
- Reference equipment and accuracy
- Use a calibrated counter, AC/DC measuring system and oscilloscope suitable for the waveform bandwidth. Record the actual termination.
- When to repeat the check
- Recheck both channels after output amplifier work. Review performance against the frequencies and amplitudes used on the bench.
Technical references 33500B Series Waveform Generators