Reference materials and instrument response
For a qualification request, share the manufacturer's performance check and the reference materials specified by your quality procedure. Keep wavelength checks, absorbance response and repeatability separately identified. A result for one lamp setting does not describe every channel on the instrument.
If your equipment record lists several element lamps, a change of lamp need not imply a change of instrument serial number. For a repeat qualification, identify the lamp and the instrument configuration used in the earlier report. That lets the reviewer compare like reference readings without treating each lamp as a separate spectrophotometer.
Atomic absorption spectrophotometer method review
Wavelength or instrument-response checks differ from verification of a complete concentration method with reference solutions. State the modules and performance evidence required. RTS checks the specified instrument functions using the analytical procedure and reference materials identified by the laboratory.
Specifications and service checks
A wavelength check can address a peak appearing at an unexpected position. A control solution that reads low raises a different question about the response of the selected analytical line. Identify the element, wavelength and observed problem so the appropriate optical or response check can be agreed.
- Range and configuration
- Agilent gives a 185–900 nm wavelength range for its cited AA instrument family. A concentration range is specific to an element, the selected absorption line and the method used.
- Manufacturers and models
- Agilent AA240FS and AA280FS are examples of atomic absorption spectrometers.
- What we check
- For wavelength indication, compare against an appropriate assigned spectral feature. A separate analytical response check uses a certified elemental reference at the concentrations relevant to the work.
- Reference equipment and accuracy
- Use suitable wavelength references and traceable elemental standards with stated concentrations and uncertainties. Record the lamp and element associated with a result.
- When to repeat the check
- Revisit the relevant checks after an optical repair or lamp change. Routine reference-material results help identify drift before the next service date.
Technical references Agilent AA instrument catalogue