On this page
Setting the anvil and measuring force
Record how the reference is set, the anvil geometry and the force setting used for the parts. List the sizes at which the bench works, including small departures from a master when it serves as a comparator.
A shop sorting a narrow band of pins can set the bench with one master at the middle of the band. A comparison at the master size and at small offsets on either side is more informative for that job than one distant scale point.
Specifications and service checks
A thin stamping that sits centrally on the anvils can give another reading near an edge if it is burred or bowed. A flat setting piece and the actual contact location help separate the bench instrument check from a part-handling question.
- Range and configuration
- As an example of the comparator heads used at a bench, the 0–25 mm Mitutoyo 510-121 has 0.001 mm graduations. Check the frame span and the shorter indicating range individually.
- Manufacturers and models
- Mitutoyo Series 510 is an example of an indicating micrometer. A dedicated bench machine can use a different head and fixture.
- What we check
- Check the spindle at several lengths, then set the comparator near zero and check small changes in both directions. Include the working stand and contacts when they affect the reading.
- Reference equipment and accuracy
- Use calibrated end standards close to the work size and known small differences between blocks. Check contact parallelism at the usual measuring force.
- When to repeat the check
- Check zero with a working master before a batch. Inspect the contacts if results begin to depend on where the part is placed.
Technical references Mitutoyo Bench micrometer reference