On this page
Tare and loading position on the platform
Record the platform size, display division, usual gross load and tare used. Include off-centre positions if operators routinely load from one side. If the head and platform can be separated, keep their identities paired for the calibration result.
When two packing stations have identical scale platforms and separate display heads, put the two serial numbers for each station on the same line. A photograph of the connected pair can settle which platform produced a result after the heads have been moved.
Specifications and service checks
A parts bag hung over the edge can touch a bench or guard and bypass part of the load. For a packing complaint, note the bag footprint and whether the same indication is obtained with the load entirely on the pan.
- Range and configuration
- An industrial bench scale and a fine laboratory balance may both display grams. Specify the plate capacity, division and normal net load. The Defender 3000 series includes 30 kg bench configurations.
- Manufacturers and models
- OHAUS Defender 3000 i-D33 is a bench-scale family. Include the indicator and platform models for a two-part scale.
- What we check
- Compare indicated weight with traceable loads at the working points. Repeatability and off-centre loading show different aspects of performance.
- Reference equipment and accuracy
- Use calibrated test weights sized for the platform and a class justified by its division and required uncertainty. Record their actual values when necessary.
- When to repeat the check
- Use the working environment and the history of checks to review recall. Moving the platform or replacing a load cell can justify calibration before the next annual date.
Technical references OHAUS Bench scale reference