Bowl speed and timing on a centrifugal extractor
If the worksheet calls for the contents to be weighed before and after extraction, the asset number on those mass readings is the balance’s. For the extractor itself, identify the type of machine, the bowl fitted, the usual running speed and any timed cycle. If the speed is selected on an ungraduated dial, tell us the position used as well as the speed your method requires.
Bitumen extractor configuration
If the machine has speed, time or temperature indications, identify the functions requiring measurement and the normal settings. Provide the equipment manual and receiving-condition requirements. The extraction result also depends on the material-testing procedure; an equipment indication check should not be described as verification of binder content.
Specifications and service checks
When a centrifuge bowl and its drive have separate tags, the speed reading belongs to the drive and the working assembly belongs to the fitted bowl. Note which bowl was installed if the enquiry follows vibration during a run.
- Range and configuration
- The B061 has a 1 litre vessel. For the heated extraction set, vessel size and the temperature measurement are relevant. Rotor speed belongs to a centrifugal model only.
- Manufacturers and models
- Matest B061 is a Kumagawa extractor. Centrifugal extractors require a different order.
- What we check
- Verify the indicators fitted to the supplied apparatus. The measured binder percentage belongs to the materials test and cannot serve as its calibration reference.
- Reference equipment and accuracy
- Use an independent temperature reference for a heater channel, or a calibrated speed reference for a centrifuge version. The weighing balance has its own calibration.
- When to repeat the check
- Review the affected channel after changing a temperature probe, drive or speed controller.