+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Bitumen penetrometer testing

A penetration reading represents movement of a needle under the load and timing conditions in the laboratory method. A dial that reads correctly can still be paired with a different needle assembly or loading weight.

Send an enquiry
Bitumen penetrometer — equipment illustration
Equipment illustration
On this page

Reading the penetration scale

On a scale graduated in tenths of a millimetre, a reading of 65 divisions represents 6.5 mm of travel. Copy the instrument’s own scale convention onto the order. A travel check alone leaves the loading weights and timed release unchecked. Tell us which of those are included and whether the conditioning bath and its thermometer are also due. Each should have an identifiable result.

Bitumen penetrometer

Give the quantities controlled by your method and identify the bath used to condition the sample. Each measurement needs its own requirement. Needle movement, applied mass and specimen temperature should not disappear into a request for a single generic penetrometer certificate.

Specifications and service checks

A needle holder can move freely through most of its travel and catch near the release position. Identify the needle assembly, release mechanism and observed travel region so a dimensional check and a release check can be requested separately.

Range and configuration
Its display resolves 0.01 mm. The needle load and release time also affect a reported penetration, even when the displacement display reads correctly.
Manufacturers and models
Matest B056-01 is a digital penetrometer. Humboldt’s H-1200 series is another laboratory penetrometer family.
What we check
Compare indicated movement against known displacements. Check the release timer and weigh the moving assembly as separate quantities.
Reference equipment and accuracy
Use gauge blocks or a calibrated displacement reference, a mass reference and a time standard. Choose the uncertainty for each quantity against its test-method limit.
When to repeat the check
Inspect the needle routinely and repeat the travel or timing checks after repair of the carriage or release.

Technical references Matest product reference · Humboldt product reference

Enquire about Bitumen penetrometer

Service enquiry