Rotor speed, timing and temperature
Tell us what speed the operator selects and whether the timed interval starts at switch-on or when the rotor reaches speed. On a refrigerated model, add the temperature check and its measuring position. The speed, time and temperature readings can then be compared with the corresponding values on your worksheet.
Blood bank centrifuge engineering measurements
List speed, time and any temperature requirements separately, using the approved equipment procedure. Refrigerated air, chamber indication and specimen-temperature conditions are not interchangeable. RTS measures speed, time and temperature under the specified engineering procedure, keeping blood-processing suitability separate from the instrument results.
Specifications and service checks
Two rotors can run at the same indicated rpm yet give different relative centrifugal forces. If stores supplies a replacement rotor, look up its working radius for the fitted bucket before carrying the old force setting into the equipment register.
- Range and configuration
- Thermo Scientific gives maximum speeds of 5,000 rpm for the Cryofuge 8 and 4,700 rpm for the Cryofuge 16. Record the installed rotor because a rotor limit may be lower.
- Manufacturers and models
- Thermo Scientific’s Cryofuge 8 and 16 are examples of refrigerated floor centrifuges.
- What we check
- Compare displayed speed with a traceable tachometer under the manufacturer’s service arrangement. Timer and refrigeration indications can be recorded as separate functions.
- Reference equipment and accuracy
- Select a tachometer rated for the rotational span and a calibrated time reference. For a temperature check, specify the measurement position and reference probe uncertainty.
- When to repeat the check
- Review following motor, rotor or speed-sensor work. Record which timer convention was used if starts and stops form part of the result.