Bypass level indicator calibration

A bypass chamber follows the liquid level in a vessel through its connecting ports. A float may drive an external magnetic flag display. The float density rating and the zero reference on the chamber affect the indicated level.

Send an enquiry
Bypass level indicator — equipment illustration
Equipment illustration
On this page

Liquid density and the level reference

Identify the liquid, density at operating temperature, tapping centres and the reference used for zero. Include any separately fitted switch or transmitter if its response also needs checking. Record which indication the operator actually uses to decide when to fill.

On a replacement chamber float, the part number matters as much as the vessel tag. Two floats of similar size can sit at different depths in the same liquid. If the flags disagree with the plant transmitter after a float change, record both indications against the same vessel level.

Bypass level indicators

Identify the chamber, float and display assembly, along with the liquid and its relevant operating conditions. Mark the reference datum and the positions at which an indication is needed. Keep any transmitter or switch fitted to the chamber separately identified. RTS measures the specified indication, transmitter output and switching levels, identifying each component in the result.

Specifications and service checks

A magnetic flap display can retain a few reversed flaps after a float moves past. Compare that pattern with the float position or another independent level observation. A local display fault and a buoyancy change will not necessarily affect the whole scale in the same way.

Range and configuration
Specify the distance between process connections and the zero datum. On a float type, liquid density can shift the float's immersion depth.
Manufacturers and models
WIKA BNA is a bypass level indicator with a magnetic float. Other level gauges include sight glasses and graduated scales.
What we check
Compare the displayed level with independently established liquid levels, noting rising and falling response. For a bare scale, verify graduation positions from the datum.
Reference equipment and accuracy
Use a calibrated height reference, with an appropriate density reference where float immersion matters. The mounting datum is part of the comparison.
When to repeat the check
Check for sticking or displaced flags after cleaning or a change in process liquid.