+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Centrifuge rotor inspection and setup

A rotor has a model, individual identifier and permitted use defined by its manufacturer. Its identity belongs beside a centrifuge speed result because the same drive can accept more than one rotor. A speed check is a measurement of the operating assembly.

Send an enquiry
Centrifuge rotor — equipment illustration
Equipment illustration
On this page

Rotor inspection and drive speed

If a separate rotor inspection is due, supply its service history and manufacturer's criteria. Keep dimensional or visual findings and measured speed in separate fields. A correct speed indication by itself says nothing about the remaining service life of the rotor.

A service entry headed ‘replaced rotor’ can obscure which result is still applicable. The last speed certificate may name the centrifuge drive, whereas a dimensional or condition report names the individual rotor. Include the previous and current rotor references when asking us to reconcile their histories.

Centrifuge rotor and the reported quantity

Rotational speed and calculated relative centrifugal force are different quantities, with the latter depending on the applicable radius. A rotor’s structural inspection and service-life records are separate from the calibration of the machine’s speed indication.

Specifications and service checks

A worksheet copied from an earlier rotor can retain the old radius after the lab changes to a different model. Record the engraved rotor code alongside the radius used by the instrument’s force display. The review can then address the configured value as well as measured speed.

Range and configuration
Capacity, geometry and speed rating belong to the precise rotor and bucket part numbers. A swing-out assembly may also require a matching set of buckets and adapters.
Manufacturers and models
Eppendorf identifies fixed-angle and swing-bucket rotors with codes such as FA-45-30-11 and A-4-81.
What we check
Record the part number, condition and service history. If a radius measurement is required for a speed-to-force calculation, define the measurement location on the specific assembly.
Reference equipment and accuracy
A calibrated dimensional instrument can support a radius measurement. The drive’s separate speed calibration does not establish a rotor’s structural condition or remaining life.
When to repeat the check
Follow the manufacturer’s inspection and retirement criteria. Inspect or refer the assembly for service after damage, corrosion or a reported incident.

Technical references Eppendorf rotor designations · Beckman Coulter Avanti rotor catalogue

Enquire about Centrifuge rotor

Service enquiry