Colorimeter testing

A colorimeter may measure the colour of a surface or the response of a liquid sample through a selected optical filter. A paint laboratory reporting colour coordinates and a laboratory reporting a solution concentration need different references and results.

Send an enquiry
Colorimeter — equipment illustration
Equipment illustration
On this page

Paint coordinates and solution readings

For a paint sample, send the colour coordinates reported by the instrument and its illuminant, observer and measurement geometry. If your laboratory uses a colorimeter for solutions, a method sheet showing the filter, cell and reported units is more useful. Both instruments may be called a colorimeter on the register, yet the references needed for the job differ.

Colorimeter: colour or solution response

Identify the illumination, viewing arrangement or selected wavelength, as applicable. Include the reference tile, filter or solution records used by your procedure. A check of one colour standard or solution should not be presented as verification of every product colour or analytical concentration.

Specifications and service checks

When two departments compare a gloss paint panel, record the finish and whether the instrument included or excluded the specular reflection. A common colour name or a matching displayed coordinate format alone cannot reconcile those settings.

Range and configuration
Konica Minolta’s CR-400 has an 8 mm measuring aperture for surface colour. A liquid filter colorimeter instead needs its filter wavelength, path length and working concentration or absorbance range.
Manufacturers and models
The Konica Minolta CR-400 Chroma Meter is a surface instrument. The sample presentation determines which checks apply.
What we check
For a surface instrument, standardise with the designated white tile and read stable colour references under fixed geometry. For a solution instrument compare absorbance using appropriate optical standards.
Reference equipment and accuracy
Use reference tiles with assigned colour coordinates for the selected illuminant and observer, or certified solution or filter standards for the analytical method.
When to repeat the check
Repeat standardisation after optical cleaning, lamp work or a change of measuring head. Inspect the reference tile for dirt and damage before use.

Technical references Konica Minolta CR-400 Chroma Meter

Enquire about Colorimeter