A compound dial covers pressures below and above atmospheric pressure. On a plant suction line the pointer may spend most of its life on the vacuum side even though most of the printed scale is positive.

Send an enquiry
Compound pressure gauge — equipment illustration
Equipment illustration
On this page

Units on either side of zero

A compound dial may use one unit for suction and another for positive pressure. Give us the unit and operating point on each side of zero. A final zero reading after the gauge has travelled through both ranges gives you a point of comparison at the next calibration.

A gauge on a reversible line might be read at minus 0.4 bar on suction and plus 2 bar on delivery. Label both as working points in the same reference system. Converting only the suction point to absolute pressure would obscure their relationship on the dial.

Specifications and service checks

Check how the instrument is used during a cleaning cycle. A line may spend most of its shift above atmospheric pressure and only cross into suction during cleaning. If operators rely on that transition, include a point close to zero on each side of it.

Range and configuration
A compound gauge spans both vacuum and positive gauge pressure. List the two full-scale limits, since they need not be equal.
Manufacturers and models
WIKA 232.50 is offered with compound scales. Check the scales printed on either side of atmospheric zero.
What we check
Check the negative side, zero crossing and positive side. If a plant uses only the vacuum sector, identify this explicitly in the order.
Reference equipment and accuracy
Use traceable references that cover both sides of atmospheric zero. A positive-pressure-only reference cannot establish the vacuum readings.
When to repeat the check
Review drift on the two sectors separately, particularly after a positive-pressure overload.

Technical references WIKA 232.50 and 233.50 Bourdon gauges

Enquire about Compound pressure gauge

Compound vacuum gauge

A suction gauge that crosses atmospheric zero can mix vacuum and positive-pressure units on the same dial. Identify the interval and the unit on each side when ordering a comparison.

Some compound dials use inches of mercury on the vacuum arc and pressure units on the positive arc. Describe a recipe point as a number with its printed unit. A photograph makes the change of scale clear to anyone entering the points on an order.

Range and configuration
A compound dial crosses atmospheric zero. Specify the minimum vacuum and maximum positive gauge pressure separately and identify the side used to release the process.

Shared with Compound pressure gauge — Manufacturers and models · What we check · Reference equipment and accuracy · When to repeat the check · Technical references

Enquire about Compound vacuum gauge

Service enquiry