Conductivity cell calibration

A conductivity cell and its associated meter can be checked as a pair, or the cell constant can be assessed for a particular probe. Identify the electrodes, their configuration and the working conductivity interval. A spare probe with a different cell constant is a separate item.

Send an enquiry
Conductivity cell — equipment illustration
Equipment illustration
On this page

The cell and its temperature compensation

If temperature compensation is active, state which temperature sensor and reference temperature are used. Retain the cell identifier alongside the meter reading so a change caused by a probe replacement can be distinguished from a change in the electronics.

Conductivity cell paired with the meter

State whether the requested result concerns the full sensor system or an electrical input simulation. A cell constant belongs to a defined cell and method; it should not be copied from another probe simply because the model is similar. Include the reference-solution conditions required by the laboratory.

Specifications and service checks

A laboratory keeps two cells on one conductivity meter, each with its own cell constant. If the spare was plugged in after cleaning, match the engraved cell ID to the constant selected on the meter. This is worth resolving before attributing the difference to the electronics.

Range and configuration
The measurable conductivity depends on the cell constant, meter range and temperature compensation. Record units carefully, as µS/cm and mS/cm differ by a factor of one thousand.
Manufacturers and models
METTLER TOLEDO’s SevenExcellence family includes S700 conductivity configurations with different sensor options.
What we check
Identify the cell separately from the meter and compare its response in an appropriate certified conductivity solution. Keep its cell constant, serial number and temperature measurement with the result.
Reference equipment and accuracy
Use traceable conductivity reference solutions stated at a reference temperature and a suitable thermometer. The certificate should identify the cell constant used.
When to repeat the check
Review the cell after cleaning, physical damage or a changed correction. Routine checks in a known solution help set the next calibration date.

Technical references METTLER TOLEDO SevenExcellence catalogue · NIST conductivity reference materials catalogue · METTLER TOLEDO conductivity cell details