On this page
Switching pressure for each contact
Identify each contact and whether the plant uses it on rising or falling pressure. A record of the operate and reset pressures can be compared with the pointer indication at the same moments. Include separate high and low alarms if both are used.
A pointer may reach its switch marker before the plant relay operates. Comparing the contact event at the gauge and the separate relay indication identifies which part of the chain is being assessed. The dial marker is not a substitute for either observation.
Specifications and service checks
A compressor may have a visible marker at 6 bar yet use only one of two available changeover contacts. Mark the terminal pair that operates the plant alarm. The pointer indication, the contact transition and the alarm set marker can then be identified separately.
- Range and configuration
- There are two quantities to specify: the pressure span of the dial and the setting of each switch contact.
- Manufacturers and models
- WIKA PGS23 combines mechanical indication with switching contacts.
- What we check
- Record indication error, the pressure at which each contact changes state and the pressure at which it resets. A dial-only result omits the switching behaviour.
- Reference equipment and accuracy
- A calibrated pressure reference measures both thresholds. A separate continuity or contact monitor records the switch state.
- When to repeat the check
- Choose the interval using the contact duty as well as dial stability. Check a contact again after its set point is adjusted.
Technical references WIKA PGS series contact pressure gauges