On this page
The differential at filter changeover
For a filter gauge, start with the pressure difference at which maintenance replaces the filter. Add the units, static line pressure and the markings on the two ports. We can include that working point directly in the report alongside any other readings you request.
A display at zero with both ports open only shows how the pointer behaves at equal pressure under that condition. If the filter change limit is 80 Pa, include a comparison near 80 Pa and identify the usual line or duct conditions.
Specifications and service checks
A filter log gives 120 Pa as a changeover point. Would staff read it while the filter is loading, or while a cleaned filter returns to service? Requesting readings on the particular side of that changeover makes the check relevant to the maintenance decision.
- Range and configuration
- Specify differential span and allowable static line pressure separately. For instance, the pressure drop across a filter can be small while both ports are under substantial line pressure.
- Manufacturers and models
- WIKA 732.14 is an example with separate high and low pressure connections.
- What we check
- Compare the differential indication as the applied difference rises and falls. State whether behaviour at elevated static pressure is part of the requested work.
- Reference equipment and accuracy
- Use a calibrated differential reference with suitable common-mode rating. Its uncertainty at the lowest requested difference is often the deciding figure.
- When to repeat the check
- Track zero drift and impulse-line incidents when choosing the next date. A new pair of process connections may change the installed result.