On this page
Choosing the base and rod for each depth
List the base, rods and display mode used for each working depth. If the tool is zeroed on a setting surface before a batch, describe the arrangement. A replacement battery or slider reset makes the original zero setting particularly important when investigating a changed reading.
When a gauge displays a positive depth on one fixture and a signed deviation on another, record the reset point used by each operator. An otherwise correct result may appear reversed or shifted if the two references are mixed in the inspection log.
Specifications and service checks
The same instrument may measure a groove with its flat tip and a drilled feature with a ball-ended tip. Mark the contact surface on the component drawing. Each tip reaches a different physical point even when the display has the same zero.
- Range and configuration
- The 0–150 mm Mitutoyo 571-201-30 displays 0.01 mm and has a stated accuracy of ±0.02 mm. The last digit and the allowed indication error are separate specifications.
- Manufacturers and models
- Mitutoyo lists both standard and hook-ended Series 571 digital depth gauges. Identify the actual measuring face used by stores or inspection.
- What we check
- With the origin set on a suitable flat, compare the measurement face against known depths. Include as-found origin and repeat readings at selected points after a fresh placement.
- Reference equipment and accuracy
- Use traceable depth references with assigned values and uncertainty small enough for the model’s error limit. Include condition and flatness of the face used to establish zero.
- When to repeat the check
- Review after a base, beam or probe repair. If the origin was lost, confirm it against a known step before using stored corrections.
Technical references Mitutoyo Digital depth gauge reference · ABSOLUTE Digimatic Depth Gage - Mitutoyo · Depth Micro Checker - Mitutoyo