Inside and outside probing directions
Identify the fitted scriber, ball probe or indicator holder. A toolroom measuring inside and outside features may use opposite probing directions. List a representative feature and height for each direction so that a single set of readings is not applied to an untested setup.
On a height setting for a turned part, a probe can approach an accessible flange from above but must approach its underside from below. A drawing showing both surfaces makes it clear why contact direction and tip compensation matter for that order.
Specifications and service checks
A workpiece base on a 10 mm parallel brings a feature 10 mm higher above the plate. If the display was zeroed on the parallel, a dimension from the plate will differ by that offset even with identical probe contact.
- Range and configuration
- The Mitutoyo 192-613-10 has a 0–300 mm range and a stated accuracy of ±0.02 mm. Identify a fitted scriber, electronic probe or indicator rather than assuming the same contact geometry for every job.
- Manufacturers and models
- Mitutoyo’s Series 192 includes electronic Digimatic and mechanical dial models. The 192-613-10 is a Digimatic example.
- What we check
- Compare indicated heights at several carriage positions using the fitted contact. Record the working datum and keep absolute and reset-origin readings identifiable where both modes are used.
- Reference equipment and accuracy
- Use a calibrated height master or blocks on a verified plate. Position the reference under the contact to limit tilt and account for plate flatness and contact repeatability.
- When to repeat the check
- A replacement scriber or a loose column changes the measuring chain. Recheck near the usual inspection heights before resuming use.
Technical references Digimatic Height Gage - Mitutoyo · Mitutoyo Height Master
Electronic height gauge
Some electronic models calculate a centre height from two opposed contacts. If that is the production measurement, include the ball tip and the probe calibration accessory rather than asking only for a height from the table.
To obtain a bore centre height, a height gauge may capture the top and bottom contacts with a calibrated probe ball. Retain both contact readings when an unexpected centre value appears, along with the probe configuration selected in the instrument.
- Range and configuration
- For one electronic example, Mitutoyo’s 192-613-10 spans 0–300 mm. Selectable display increments elsewhere in Series 192 depend on the model and should be recorded with the readings.
- Manufacturers and models
- Mitutoyo’s electronic Digimatic Series 192 instruments illustrate why a complete model suffix is needed on the order.
- What we check
- Check repeatability at the datum, then compare indicated height with known steps over the working travel. If the gauge accepts a removable probe, identify the actual probe and stylus used.
- Reference equipment and accuracy
- Use a calibrated height standard or suitable gauge blocks on a flat plate of known grade. Both the reference-height uncertainty and contact deflection matter on a fine digital display.
- When to repeat the check
- Review results after a probe change, encoder service or evidence that zeros shift between parts. A single zero check does not test the full working travel.
Technical references Mitutoyo Digimatic Height Gage · Digimatic Height Gage - Mitutoyo