On this page
Spindle setup and the indicator’s data output
Record the contact tip, spindle orientation and display mode used on the stand. If a production system reads the data port, say whether that output needs to be compared as well. For a tolerance centred on zero, request displacement points on each side of the operating position.
If a production fixture expects the display to decrease when a pin protrudes, the operator may have reversed the counting direction. Record which direction gives a positive value before comparing the display with values downloaded from the inspection system.
Specifications and service checks
A fixture may preset the indicator to the nominal part dimension instead of zero. Record the preset when saving a reading. Otherwise a value such as 12.008 mm can be mistaken for a displacement of 12 mm rather than a small size departure.
- Range and configuration
- The Mitutoyo 543-310B documentation lists selectable 0.001 to 0.01 mm resolution. Display increment, spindle stroke and maximum permissible indication error are separate entries.
- Manufacturers and models
- Mitutoyo’s ABSOLUTE Digimatic indicators include 543-series instruments. Record the complete suffix because spindle travel and available display modes vary.
- What we check
- Compare displayed displacement with an independent reference along the spindle. Record the units, preset and direction setting. A reversed display may be intentional in the installed fixture.
- Reference equipment and accuracy
- Use a calibrated indicator tester or traceable steps in a rigid comparator stand. At the finer setting check that reference uncertainty and stand repeatability support the requested tolerance.
- When to repeat the check
- Recheck after a tip or mounting-stem change. Trend a mid-travel point as well as the datum when the gauge is used as part of a shop comparator.
Technical references Mitutoyo ABSOLUTE Digimatic Indicator ID-C · Mitutoyo ABSOLUTE Digimatic Indicator ID-C