On this page
Zero reference and the micrometer’s data output
Identify the range, display unit and reference used when zeroing. If the inspection system records data through a cable, include that output in the request. Check points should span the spindle positions used for the parts, including intermediate screw positions.
When a micrometer is set with a reference rod at the start of a shift, record that rod and the preset value. A later comparison can then distinguish a wrong preset from a change in the spindle or anvil faces.
Specifications and service checks
An operator can leave a digital micrometer in inch mode while a drawing is in millimetres. A display of 0.5000 in represents 12.7 mm. Keep the unit in the exported inspection record so the digits remain meaningful.
- Range and configuration
- A digital reading of 0.001 mm is common on shop instruments. The Mitutoyo 293-100-10 is a finer 0–25 mm example with a stated accuracy of ±0.5 µm under its specified conditions.
- Manufacturers and models
- Mitutoyo’s Series 293 includes models with different spans and resolutions. The full part number distinguishes the high-resolution MDH from other digital micrometers.
- What we check
- Compare the displayed length with calibrated blocks at several spindle positions. Record the display increment and inspect whether the result changes with the contact position on the faces.
- Reference equipment and accuracy
- Select end standards by their calibration uncertainty and condition. At sub-micrometre limits, reference corrections, temperature and measuring force can outweigh the last displayed digit.
- When to repeat the check
- Retain a check near the usual work size as well as the daily zero. Review the interval after measuring-face work or a change in the repeatability of the force mechanism.
Technical references MDH Micrometer - Mitutoyo · Digimatic Micrometer - Mitutoyo