+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Digital micrometer calibration

A digital micrometer may display an absolute spindle position or a difference from a reference piece. A preset for an extended measuring range must agree with the anvils fitted. The electronic value and the engraved sleeve reading can also be separate outputs on some models.

Send an enquiry
Digital micrometer — equipment illustration
Equipment illustration
On this page

Zero reference and the micrometer’s data output

Identify the range, display unit and reference used when zeroing. If the inspection system records data through a cable, include that output in the request. Check points should span the spindle positions used for the parts, including intermediate screw positions.

When a micrometer is set with a reference rod at the start of a shift, record that rod and the preset value. A later comparison can then distinguish a wrong preset from a change in the spindle or anvil faces.

Specifications and service checks

An operator can leave a digital micrometer in inch mode while a drawing is in millimetres. A display of 0.5000 in represents 12.7 mm. Keep the unit in the exported inspection record so the digits remain meaningful.

Range and configuration
A digital reading of 0.001 mm is common on shop instruments. The Mitutoyo 293-100-10 is a finer 0–25 mm example with a stated accuracy of ±0.5 µm under its specified conditions.
Manufacturers and models
Mitutoyo’s Series 293 includes models with different spans and resolutions. The full part number distinguishes the high-resolution MDH from other digital micrometers.
What we check
Compare the displayed length with calibrated blocks at several spindle positions. Record the display increment and inspect whether the result changes with the contact position on the faces.
Reference equipment and accuracy
Select end standards by their calibration uncertainty and condition. At sub-micrometre limits, reference corrections, temperature and measuring force can outweigh the last displayed digit.
When to repeat the check
Retain a check near the usual work size as well as the daily zero. Review the interval after measuring-face work or a change in the repeatability of the force mechanism.

Technical references MDH Micrometer - Mitutoyo · Digimatic Micrometer - Mitutoyo

Enquire about Digital micrometer

Service enquiry