Live, logged and peak pressure readings
Send the unit, pressure reference, display resolution and any tare in use. If the issue appeared in a logged or peak-hold value, tell us whether the steady display showed the same difference. A detachable sensor and its display should be identified together.
On a plant where a gauge stores a peak while the operator reads the live display, retain both field names. An apparent disagreement during a changing pressure cycle is not automatically an indication error at a steady calibration point.
Specifications and service checks
A technician switching a gauge from bar to MPa should see a 5 bar reading become 0.5 MPa. If the electronic log still exports the numeral 5, check the unit attached to the saved field. A screenshot and an exported row are useful when the complaint appears only after downloading.
- Range and configuration
- The WIKA CPG1500 series includes gauge-pressure ranges up to 1,000 bar and different accuracy options. Confirm the individual full scale and whether the permitted error is specified as a percentage of span or reading.
- Manufacturers and models
- WIKA CPG1500 is a digital reference gauge available in several accuracy options. Mechanical test gauges have a dial class instead.
- What we check
- Compare the working points on increasing and decreasing pressure and record the display unit and zero setting. Retain the as-found readings before any correction.
- Reference equipment and accuracy
- Use a reference controller or pressure balance whose uncertainty resolves the required limit. More display digits alone do not establish reference quality.
- When to repeat the check
- For a gauge used as a master, review the calibration history and intermediate comparisons with an independent check gauge.
Technical references CPG1500 Precision Digital Pressure Gauge - WIKA USA