Digital thermometer calibration

A digital thermometer may display more decimal places than its probe and electronics can measure accurately. Give the working points and the allowed error in your process, along with the resolution. A request based only on the number of decimal places leaves the requirement unclear.

Send an enquiry
Digital thermometer — equipment illustration
Equipment illustration
On this page

Which probe uses which socket?

On multi-input models, identify the socket and paired probe for each result. If your team moves probes between sockets, tell us which combinations it uses and keep each pairing on the equipment register.

A thermometer reads 100.4 °C when the reference reads 100.0 °C. In that example the indicated error is +0.4 °C, while the correction to add to the indication is −0.4 °C. Keep the certificate's own sign convention with the entry in the asset register.

Specifications and service checks

A spare probe can fit the socket and still have a different response curve. If a step change followed a probe exchange, send the probe code and meter input setting with the asset ID. They establish which combination the operator actually reads.

Range and configuration
Fluke lists −200 to +1,372 °C for type K and −250 to +400 °C for type T on the 54 II B. Those figures describe the input electronics. The attached probe and lead can have a narrower rating.
Manufacturers and models
Fluke’s 54 II B is a dual-input thermocouple thermometer. Its 1523 reference readout also accepts suitable resistance and thermistor probes.
What we check
For a two-input instrument, pair each result with its input and thermocouple type. An electrical simulation tests the display and input chain. Include a comparison with the fitted probe when the instrument is used to read temperatures directly.
Reference equipment and accuracy
Use a traceable reference thermometer with current corrections. Select the source and immersion depth for the sensor geometry and the uncertainty required.
When to repeat the check
Record whether the previous result applied to this exact probe. A replacement sensor or an impact can bring the next complete-system check forward.

Technical references Fluke 54 II thermometer · Fluke 1523 and 1524 guide · BIPM guide to industrial platinum resistance thermometers