+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Disc micrometer calibration

Disc anvils present thin, broad contact faces that can reach across gear teeth or onto a recessed shoulder. The diameter and thickness of the discs determine where contact occurs. A normal flat-anvil micrometer does not reach the same feature.

Send an enquiry
Disc micrometer — equipment illustration
Equipment illustration
On this page

Where the measuring discs touch the work

Include the fitted discs and mark the region used on a typical part. Give the span or working dimension and any requirement for checks across the disc faces. A result taken near the centre may not describe contact near a disc edge.

For a measurement over several gear teeth, show the number of teeth spanned and the face width. The purpose is to identify the position and opening of the discs, independently of the tooth-geometry inspection itself.

Specifications and service checks

When checking the span across gear teeth, bearing on the chamfer instead of the working flanks can shift the result. A drawing with the measured span and the permitted contact region helps select which portion of the discs to examine.

Range and configuration
Mitutoyo lists measuring discs 20 mm in diameter for its Series 323, 223 and 123 rotating-spindle families. Disc diameter and edge thickness determine access, while the screw reading gives the measured length.
Manufacturers and models
Mitutoyo’s 123-125 is an example of a mechanical disc micrometer. Its broad measuring contacts are different from conventional flat cylindrical anvils.
What we check
Compare indicated length with end standards at several readings. Check how consistently the opposing discs meet across their working areas. Thin edges can show damage before the centre reading changes.
Reference equipment and accuracy
Use calibrated length standards and a method that resolves disc parallelism without bending the edges. Select reference uncertainty against the allowed span measurement error.
When to repeat the check
Inspect after an edge strikes a tooth or groove. Recheck any face that has been dressed and keep the working disc condition with the results.

Technical references Mitutoyo Disc micrometer reference · Mitutoyo Disk Micrometers

Enquire about Disc micrometer

Service enquiry