Stroke and output at the receiving instrument
List the stroke used, the output signal and the readout or controller that converts it to length. For a non-contact head, identify the normal target surface. For a spring-loaded probe, include the fitted tip and how far it is pressed in at the reference position.
A probe used to follow a press stroke may be read in the controller as millimetres from the closed position, while the transducer increases its output toward the open position. Recording both end positions prevents a reversed scale from being mistaken for a sensor error.
Specifications and service checks
A 0–20 mm sensor scaled linearly to 0–10 V would ideally give 5 V halfway through travel. A different midpoint on the plant screen can come from the sensor or the input scaling. Keep the electrical and displayed values together.
- Range and configuration
- Give the active sensing stroke and electrical output span. A sensor operating between two limits on a press may use only a fraction of its physical travel.
- Manufacturers and models
- Solartron Metrology’s G-Type is an LVDT-based probe family with conditioning electronics.
- What we check
- Move the probe through traceable displacements in both directions. Record its output and the excitation or conditioner settings, checking zero, slope and return error.
- Reference equipment and accuracy
- Use a displacement comparator, micrometer stage or laser system with assigned uncertainty. An electrical reference is also needed when the output is being calibrated in volts.
- When to repeat the check
- Recheck after changing the conditioner gain or probe mounting. Track zero and a mid-stroke check between full calibrations.