Displacement sensor calibration

A displacement sensor might read a moving contact, a target across an air gap or the position of a sliding core. The target material, alignment and zero position can all be part of the measurement. Determine which movement the installed system is intended to report.

Send an enquiry
Displacement sensor — equipment illustration
Equipment illustration
On this page

Stroke and output at the receiving instrument

List the stroke used, the output signal and the readout or controller that converts it to length. For a non-contact head, identify the normal target surface. For a spring-loaded probe, include the fitted tip and how far it is pressed in at the reference position.

A probe used to follow a press stroke may be read in the controller as millimetres from the closed position, while the transducer increases its output toward the open position. Recording both end positions prevents a reversed scale from being mistaken for a sensor error.

Specifications and service checks

A 0–20 mm sensor scaled linearly to 0–10 V would ideally give 5 V halfway through travel. A different midpoint on the plant screen can come from the sensor or the input scaling. Keep the electrical and displayed values together.

Range and configuration
Give the active sensing stroke and electrical output span. A sensor operating between two limits on a press may use only a fraction of its physical travel.
Manufacturers and models
Solartron Metrology’s G-Type is an LVDT-based probe family with conditioning electronics.
What we check
Move the probe through traceable displacements in both directions. Record its output and the excitation or conditioner settings, checking zero, slope and return error.
Reference equipment and accuracy
Use a displacement comparator, micrometer stage or laser system with assigned uncertainty. An electrical reference is also needed when the output is being calibrated in volts.
When to repeat the check
Recheck after changing the conditioner gain or probe mounting. Track zero and a mid-stroke check between full calibrations.