Tie travel to a reference position
Identify the sensor type, stroke and contact tip or moving target. On an LVDT, state whether readings are taken on either side of the electrical null or from a zero set on the display. Mark which direction the machine calls positive. A sketch of the fixture can be clearer than a list of signed readings.
Include the part that converts travel to a reading
Some sensors plug into a dedicated conditioner and show millimetres on a panel. Others return a voltage or current to a controller. Tell us which output is used and whether you need the sensor on its own or the sensor and indicator as a pair. For a non-contact sensor, describe the normal target material and mounting gap.
When the fixture travels in both directions
Consider a fixture that checks the same shoulder on approach and again as it retracts. Put both readings against the same nominal position, labelled by direction. An average would hide any difference between the strokes. Note where the fixture was zeroed and which conditioner or display was connected so a technician can repeat the comparison.
Equipment covered
Linear displacement transducer
A system zeroed near mid-stroke can report movement on either side of its reference. Give both directions of travel and identify the signal conditioner used to turn the transducer output into units of length.
With the core positioned at its electrical null, equal travel in opposite directions can produce opposite-signed outputs. Retain the polarity of each reading when the instrument is used to centre a slide or check travel on both sides of zero.
- Range and configuration
- A linear probe may be specified by a total travel or by a positive and negative range about electrical zero. Give the working stroke and whether the output is voltage, ratio or a digital length.
- Manufacturers and models
- Solartron Metrology’s G-Type is an LVDT-based probe family with conditioning electronics.
- What we check
- Move the probe through traceable displacements in both directions. Record its output and the excitation or conditioner settings, checking zero, slope and return error.
- Reference equipment and accuracy
- Use a displacement comparator, micrometer stage or laser system with assigned uncertainty. An electrical reference is also needed when the output is being calibrated in volts.
- When to repeat the check
- Recheck after changing the conditioner gain or probe mounting. Track zero and a mid-stroke check between full calibrations.
Technical references Solartron Metrology Microsoft PowerPoint - 503144 G-Type update