+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Dry film thickness gauge calibration

A dry film thickness reading normally comes from a probe resting on a cured coating over its substrate. Probes intended for ferrous and non-ferrous bases can use different measuring principles. A dual-mode gauge may require results on both bases.

Send an enquiry
Dry film thickness gauge — equipment illustration
Equipment illustration
On this page

Probe and substrate beneath the coating

Identify the substrate, probe and typical coating thickness. If you use foils on a bare part for the zero and span checks, keep the foil values and base material with the results. Rough or curved surfaces can behave differently from a smooth flat plate.

A probe zeroed on flat bare steel may behave differently on a small coated tube. If tubes are the parts being accepted, include their material and diameter, and indicate whether an uncoated piece of the same shape is available for the comparison.

Specifications and service checks

On a blasted steel panel, several readings over a small patch can vary as the probe crosses peaks and valleys. Mark the patch on a sketch and retain the individual readings if the disagreement is with another inspector’s average.

Range and configuration
An Elcometer 456 Scale 1 integral gauge covers 0–1,500 µm. The probe and base material determine the usable range on steel or non-ferrous metal.
Manufacturers and models
Elcometer 456 is available with different probe and scale options. Mextech also supplies CM-series coating thickness gauges.
What we check
Check zero on an appropriate bare substrate, then compare readings on assigned coating standards or foils over the working thicknesses. Include curvature or roughness if it changes the application.
Reference equipment and accuracy
Use certified dry-film standards or assigned foils on the intended base metal. For a rough blasted surface, agree how base-metal readings are treated before evaluating thin coats.
When to repeat the check
Check the response on a known thickness before inspection and after a probe change. A worn probe tip can give different results on flat and curved parts.

Technical references Elcometer 456 Integral Coating Thickness Gauge · CC-4643 calibration scope, page 15

Enquire about Dry film thickness gauge

Service enquiry