Electronic balance calibration

An electronic balance can have internal adjustment, selectable filtering and more than one display range. The state of those settings matters when comparing readings from different days. A startup adjustment message is distinct from a comparison with identified external reference weights.

Send an enquiry
Electronic balance — equipment illustration
Equipment illustration
On this page

Gross load and the smallest display increment

List the normal range, displayed increment and the point at which a second range or division becomes active. State whether received-condition readings are needed before adjustment. If results are recorded remotely, identify the displayed and transmitted units used for routine weighing.

A balance may change its displayed increment above a range boundary. If two product sizes fall on opposite sides, include points in both ranges. The report can then preserve the resolution actually used for each product.

Specifications and service checks

A printer may round more heavily than the balance display. Compare the printed label and the display for one unchanged load before reporting a discrepancy. Record each unit and the number of decimal places transferred to the batch record.

Range and configuration
For site calibration, entry 83 of the RTS scope covers an electronic balance from 0 to 220 g with 0.001 g readability, accuracy class I or coarser. The listed comparison uses E2 weights.
Manufacturers and models
OHAUS Pioneer and Defender instruments serve different load ranges. Supply the exact code, capacity and readability so the comparison can be matched to the specific balance.
What we check
Check indicated mass with traceable loads, then move a repeatable load over the pan to assess eccentric loading. If a large tare is used, include the net weighing region that remains available.
Reference equipment and accuracy
Use calibrated test weights sized for the platform and a class justified by its division and required uncertainty. Record their actual values when necessary.
When to repeat the check
Use the working environment and the history of checks to review recall. Moving the platform or replacing a load cell can justify calibration before the next annual date.