Menu

Electronic pipette calibration

An electronic pipette can retain distinct operating modes for an ordinary transfer, repeated dispensing or a programmed series. A volume check made in one mode should be traceable to that mode. Include the tip type and the settings used in the laboratory.

Send an enquiry
Electronic pipette — equipment illustration
Equipment illustration
On this page

Which dispensing programme is used?

If several programmed volumes are required, supply their names or values and identify the result required for each. Retain the software configuration or programme identifier with the report so a later device reset can be recognised in the service history.

For a pipette routinely set to make a series of equal deliveries from one aspiration, the first and last dispenses can be identified separately in the results. A comparison run in single-transfer mode would not directly check the sequence saved under that repeating programme.

Specifications and service checks

In repeat-dispense mode, one aspiration supplies several deliveries. If only the last aliquot causes concern, list the programmed series and identify its position in that series. A comparison at the same nominal volume in single-delivery mode would answer a different question.

Range and configuration
Eppendorf’s Xplorer series spans 0.5 µL to 10 mL across different models. Identify the volume on the particular head and the mode used for reporting a delivered amount.
Manufacturers and models
Eppendorf’s Xplorer and Xplorer plus are electronic piston pipettes. Gilson publishes model- and mode-specific data for PIPETMAN M.
What we check
Compare deliveries for the identified channels and modes, preserving channel order on the report. A mean across all channels can hide an error at one tip position.
Reference equipment and accuracy
Use a traceable gravimetric arrangement or another applicable validated volume method. Its uncertainty and resolution must support one channel’s smallest delivery.
When to repeat the check
Review after a head or seal change and after persistent channel-to-channel differences. Include the channel identity on interim checks.

Technical references Eppendorf Xplorer manual · Gilson PIPETMAN M specifications · ISO 8655-6 gravimetric volume determination