Electrosurgical unit testing

An electrosurgical generator can have several output modes with the same power setting on the panel. The matching bench record identifies the mode, test load and output connector alongside the measured value.

Send an enquiry
Electrosurgical unit — equipment illustration
Equipment illustration
On this page

Generator modes and footswitch identification

Include the installed options and the names of the modes in the engineering procedure. A footswitch or hand control may belong to a particular theatre set and carry its own service history. Identify those items separately from the generator and retain the analyser reference with the result.

For example, a workshop may record a value at a named setting and later compare it with the same setting in a different mode. Include the mode and load on both lines. Equal numbers on the front panel do not necessarily mean the bench setup or expected output was the same.

Specifications and service checks

When two workshop records show different power readings, compare the selected mode and analyser load first. A wattage value copied without the load or output port is difficult to interpret. Include those details from the engineering setup.

Range and configuration
RF output is specified by mode and test load. A single wattage reading cannot describe every output configuration.
Manufacturers and models
Erbe provides a biomedical verification document for its VIO 3 electrosurgical unit.
What we check
A trained engineering team measures delivered power with an appropriate ESU analyser under the model-specific verification procedure. Identify the mode, output and reference load on the report.
Reference equipment and accuracy
Use an analyser calibrated for the required RF characteristics, load and power. A low-frequency bench power meter has a different measurement bandwidth.
When to repeat the check
Review after generator repairs, connector replacement or service software changes. The hospital’s programme should track the release of each output checked.

Technical references Erbe product reference