On this page
Working distance between pins
For an elongation gauge with several pin pairs, label the pair by aggregate fraction and mark the working gap on the drawing. On a worn pair, a result taken at the base of a pin may differ from one nearer the top. Tell us where the method defines the working dimension. That gives the materials laboratory a result it can relate to the particles it classifies.
Elongation gauge openings
Send the reference dimensions from the applicable method, rather than a single overall board size. Describe loose pins or changes in the supporting base. RTS measures the specified gauge features against the identified dimensional requirement.
Specifications and service checks
A pin that has bent inwards changes one opening without affecting the other size positions. Mark the affected station with its size fraction so the lab can check the working gap the operator actually uses.
- Range and configuration
- The pin gaps represent aggregate size fractions. The Matest design classifies elongated particles using a length of 1.8 times nominal size.
- Manufacturers and models
- Matest A071 is a fixed-pin length gauge for aggregate.
- What we check
- Measure each working gap at the contact height and inspect pin alignment. Record each fraction separately so one damaged pin pair can be identified.
- Reference equipment and accuracy
- Use calibrated length standards or a suitable inside measuring instrument. Select uncertainty against the permitted pin-gap error.
- When to repeat the check
- Inspect for loose pins before use and remeasure any pair that has shifted or suffered impact.
Technical references Matest product reference