Energy calibrator calibration

An energy calibrator may supply voltage and current with a controlled phase relationship, compare meter pulses, or perform both tasks. Its power output and its time or pulse reference contribute to an energy comparison.

Send an enquiry
Energy calibrator — equipment illustration
Equipment illustration
On this page

Matching pulse constants on a meter bench

For a meter marked 1,000 imp/kWh, each counted pulse represents 1 Wh. If a bench is configured for a different pulse constant, its calculated energy will disagree even with steady voltage and current. Give the constant and the pulse input for each position in use, then the test voltage, current, frequency, phase arrangement and leading or lagging power factor. This ties the energy result to the actual meter routine.

Energy calibrators that generate a meter test

Identify the source channels, phase arrangement, pulse interface and reference energy function. State whether your result is expressed as power, energy or meter error. Include frequency and phase conditions for each requested point. Record the reference direction and complete measurement arrangement so the generated outputs and energy comparison are evaluated together.

Specifications and service checks

For a three-phase meter bench, total active energy can depend on the phase angles set between individual voltage and current channels. A balanced unity-power-factor point and a low-power-factor point are distinct requests even at the same nominal voltage and current.

Range and configuration
Specify volts, amperes, power factor and the integration interval. At 1 kW for 6 minutes the nominal active energy is 0.1 kWh.
Manufacturers and models
Fluke 6105A is an electrical power standard used for energy-meter calibration. Its energy output involves voltage, current, phase and time.
What we check
Compare delivered energy with an independent standard; where relevant verify the pulse output and pulse count against the accumulated quantity.
Reference equipment and accuracy
Use a calibrated reference energy meter or power comparator with traceable timing. Phase accuracy matters particularly at low power factor.
When to repeat the check
Monitor reference loads and reassess after repairs to a voltage, current or timing channel.

Technical references Using the 6105A Electrical Power Standard to calibrate energy meters