On this page
The marked volume and tip system
Identify the marked volume, tip system and any service work completed before the check. If as-found results are required, request them before adjustment. For a set of identical pipettes, mark each body so its result follows it back to the correct bench.
Two 100 µL pipettes on one order may need different handling if one is a working instrument due for routine service and the other is being investigated after a drop. Mark the second by serial number and request its initial delivery results before any adjustment.
Specifications and service checks
A bench has two fixed-volume pipettes marked 100 µL, one issued with a different tip system. Keep the two asset numbers and tip references separate for a delivery comparison. The visible volume marking gives the nominal setting on both handles, even when their individual deliveries differ.
- Range and configuration
- Socorex’s fixed-volume Acura families include nominal volumes from 1 µL to 10 mL across the range. Each individual pipette delivers the particular volume marked on its body.
- Manufacturers and models
- Socorex’s Acura manual 815 and 835F are fixed-volume instrument families.
- What we check
- Measure repeated deliveries with the specified tip and report mean delivery and repeatability at the marked volume.
- Reference equipment and accuracy
- Use a calibrated balance appropriate to the mass of one delivery, a temperature measurement and evaporation control as needed. The uncertainty should support the permitted volume error.
- When to repeat the check
- Review results from routine delivery checks. A replaced piston, seal or different approved tip calls for confirmation at nominal volume.
Technical references Socorex laboratory instruments catalogue · ISO 8655-6 gravimetric volume determination