Flash point apparatus testing

Before quoting a flash point apparatus check, we look at the cup design, temperature sensor and test method. An open-cup and a closed-cup instrument can carry the same temperature range but require different checks.

Send an enquiry
Flash point apparatus — equipment illustration
Equipment illustration
On this page

Instrument readings and reference material

List the controller readings, timing or programmed heating functions your method specifies. If you also need a complete run with a certified reference material, include its identity and assigned value. We can agree how to report that result alongside the temperature or timing checks.

On an older bench unit the thermometer may be removed for calibration and returned to a different cup assembly. Put the thermometer identifier and cup type on the service order. If the laboratory also holds results for the complete apparatus, the reviewer can see which thermometer was fitted for each run.

Flash point apparatus method

Identify the temperature sensor, timing functions and detection arrangement that require review. RTS checks the apparatus under the specified operating conditions, recording the selected measurements and references. A temperature comparison by itself is not a flash-point determination or method validation.

Specifications and service checks

A method name on a worksheet determines which cup and which instrument readings need attention. Identify open- or closed-cup configuration and the fitted thermometer against the named method before comparing two instruments.

Range and configuration
The manufacturer lists 40–360 °C for this configuration. Confirm cup type and thermometer designation when specifying a check.
Manufacturers and models
Matest B094A is a Pensky–Martens closed-cup tester.
What we check
Compare the temperature indication with a suitable reference. A method-specific certified reference material can assess overall apparatus performance as a separate result.
Reference equipment and accuracy
Use a traceable thermometer over the requested range. Match any certified material to the prescribed test method and its stated value and uncertainty.
When to repeat the check
Follow the applicable test method for verification frequency. Repeat checks after changes to the probe or temperature control.

Technical references Matest product reference