+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Flow indicator calibration

A panel can display flow without a fluid connection of its own. It may receive a current, frequency or pulse signal from a field instrument. The input type and configured scale determine the flow rate shown on the panel.

Send an enquiry
Flow indicator — equipment illustration
Equipment illustration
On this page

Converting an input signal to displayed flow

Include the input type, zero and full-scale flow values, and the time unit on the display. If the indicator also accumulates a total, identify its separate pulse or integration settings. An input-to-display comparison and a full meter run can then be specified separately.

A line moving at 30 litres per minute for ten minutes delivers 300 litres at that steady rate. The rate display and the totaliser can therefore be checked against two different expected values. Mark the time basis used on the panel, especially if the production sheet records hourly flow.

Flow indicators and scaling

Give the input type, displayed units, scaling and any square-root or totalising function. State whether you need an input simulation or a comparison using actual flow through the connected meter. Those results have different boundaries. Keep the sensor identity and configured range with the indicator so the enquiry does not treat a display check as a complete flow-system calibration.

Specifications and service checks

After a display replacement, the instantaneous rate may look right while the plant total builds too quickly. Identify which pulse or analogue input drives each function. A rate indication and the count accumulated over a shift may be configured independently.

Range and configuration
A float scale identifies the fluid and reference conditions. An electronic repeater marked in the same units needs its input scaling recorded instead.
Manufacturers and models
KROHNE DK32 and DK34 are local variable-area indicators for low liquid or gas flow.
What we check
Compare a float reading at steady flows against the reference. For a repeater, compare the displayed engineering units with signals representing the scale endpoints and working point.
Reference equipment and accuracy
Use a flow reference for the complete indicator or a traceable signal source for the repeater alone.
When to repeat the check
Check for float sticking and scale changes. Reverify a repeater after its scaling is edited.

Technical references DK32/34 For low liquid and gas flows

Enquire about Flow indicator

Service enquiry