Flow switch calibration

A flow switch operates a contact once the flow reaches a set condition. The rate at which it closes on increasing flow can differ from the rate at which it opens again as flow falls.

Send an enquiry
Flow switch — equipment illustration
Equipment illustration
On this page

Switching flow in the installed pipe

For a flow switch, give the fluid, pipe size, mounting position and contact the plant uses. List the trip flow and reset flow in the working units, with the direction of change for each. Include the cooling-flow threshold if that contact is tied to a machine start.

A cooling circuit may start an alarm as flow falls but reset only when the flow rises past a second value. A maintenance note of “trips at 8” is ambiguous without the flow unit and direction. Record the actual contact used by the alarm controller as well as the switch tag.

Flow switch operating points

State whether both operation and reset are to be observed and identify the contact or output used. A continuity check confirms a contact state, not the flow at which it changes. RTS measures the operating and reset flow under the stated medium, orientation and connection conditions.

Specifications and service checks

For a lubrication circuit that must prove circulation before a machine starts, the switch’s rising-flow point matters at startup. The point where the machine trips on loss of flow is approached from the opposite direction. Give each plant action its own line in the enquiry.

Range and configuration
Specify the actual trip and reset requirements in flow or velocity units, plus direction of approach. A percentage on an indicator may be relative to a taught span.
Manufacturers and models
The ifm SI5000 monitors flow and provides a switching output. A velocity setting on this type needs the pipe and medium identified.
What we check
Vary flow through the switching point from either side and record where the output changes state. The difference is the switching hysteresis.
Reference equipment and accuracy
Use a calibrated flow reference for the medium and an output-state monitor. For velocity, include the effective cross-section in the result.
When to repeat the check
Repeat after relearning thresholds or changing the pipe insertion depth.

Technical references SI5000 - Flow monitor - ifm

Enquire about Flow switch