Reading the defrost cycle on the temperature trace
Include the normal control setting and any automatic defrost period in your request. When reviewing a trace, note whether a peak coincides with defrost, access or an unrelated interruption. This makes the reported extremes easier to interpret.
For a chest freezer used across several shifts, the monitoring probe may be clipped near the lid while stock occupies several levels below it. Mark both the probe and the normal loading height. That gives the quality team a practical way to interpret the mapping positions.
Specifications and service checks
A min/max display retains an excursion after the cabinet has returned to its normal condition. Before resetting it, record all three values and the time of the last reset. This distinguishes a historical high from a present disagreement with the storage logger.
- Range and configuration
- Liebherr gives −10 to −26 °C for the LGPv 1420 laboratory freezer. Ultra-low freezers are a different equipment class and require references rated for their own storage temperatures.
- Manufacturers and models
- Liebherr’s LGPv range and Eppendorf’s CryoCube F740hi illustrate different freezer classes.
- What we check
- Compare the unit’s indication with independent readings in the storage space. Record shelf position, loading and any defrost or door events during the agreed observation.
- Reference equipment and accuracy
- Use probes whose calibration extends to the lowest temperature required. Lead routing and settling time can affect readings, especially in an ultra-low unit.
- When to repeat the check
- Check again after a refrigeration repair, control probe replacement or persistent monitor alarm. Compare drift near the normal storage temperature.
Technical references Liebherr laboratory freezer manual · Eppendorf CryoCube F740hi manual