Waveform, amplitude and receiving impedance
Specify sine, square or other waveform, output frequency, amplitude convention, DC offset and actual termination. If a square wave drives a timer, include pulse width or duty cycle as used. Identify which output channel and sync connection belong to the job.
On a sine wave with no DC offset, 2 V peak-to-peak is about 0.707 V RMS. The same conversion does not apply to a square wave. If one bench sheet uses RMS and another uses peak-to-peak, keep the waveform with the figure as well as the termination.
Specifications and service checks
A duty-cycle control may be used for square waves even when the bench usually selects sine waves. If an actuator test needs a 2 ms pulse every 10 ms, that corresponds to 20% duty. Quote the pulse requirement as well as the repetition rate.
- Range and configuration
- Amplitude can be stated for a matched 50 Ω load or a high-impedance termination. The indicated and observed voltages can differ by a factor of two if those settings disagree.
- Manufacturers and models
- Keysight 33500B-series generators include 20 MHz and 30 MHz models. Identify the output channel and waveform mode.
- What we check
- Compare frequency, amplitude and DC offset at selected waveforms and loads. For an arbitrary waveform, also identify sample rate and the features to be checked.
- Reference equipment and accuracy
- Use a calibrated counter, AC/DC measuring system and oscilloscope suitable for the waveform bandwidth. Record the actual termination.
- When to repeat the check
- Recheck both channels after output amplifier work. Review performance against the frequencies and amplitudes used on the bench.