Function generator calibration

Before comparing a function generator with a scope, check what the amplitude figure on the generator means. On some models the display shows the voltage expected across a selected load. It also matters whether the figure is peak, peak-to-peak or RMS.

Send an enquiry
Function generator — equipment illustration
Equipment illustration
On this page

Waveform, amplitude and receiving impedance

Specify sine, square or other waveform, output frequency, amplitude convention, DC offset and actual termination. If a square wave drives a timer, include pulse width or duty cycle as used. Identify which output channel and sync connection belong to the job.

On a sine wave with no DC offset, 2 V peak-to-peak is about 0.707 V RMS. The same conversion does not apply to a square wave. If one bench sheet uses RMS and another uses peak-to-peak, keep the waveform with the figure as well as the termination.

Specifications and service checks

A duty-cycle control may be used for square waves even when the bench usually selects sine waves. If an actuator test needs a 2 ms pulse every 10 ms, that corresponds to 20% duty. Quote the pulse requirement as well as the repetition rate.

Range and configuration
Amplitude can be stated for a matched 50 Ω load or a high-impedance termination. The indicated and observed voltages can differ by a factor of two if those settings disagree.
Manufacturers and models
Keysight 33500B-series generators include 20 MHz and 30 MHz models. Identify the output channel and waveform mode.
What we check
Compare frequency, amplitude and DC offset at selected waveforms and loads. For an arbitrary waveform, also identify sample rate and the features to be checked.
Reference equipment and accuracy
Use a calibrated counter, AC/DC measuring system and oscilloscope suitable for the waveform bandwidth. Record the actual termination.
When to repeat the check
Recheck both channels after output amplifier work. Review performance against the frequencies and amplitudes used on the bench.