Gas conditions behind the stated flow
Give the gas or mixture, inlet and outlet conditions, and the base temperature and pressure used for standard volume. Identify any gas correction selected in the meter. If one instrument is moved between gases, list the configurations for which results are required.
If a flow sheet states 20 standard L/min, look up the temperature and pressure represented by “standard” on that sheet and in the instrument settings. Two configurations can describe the same gas flow with different numerical values when they use different reference temperatures.
Specifications and service checks
A calibration gas and the gas used on the line may differ. Identify both, together with the meter principle. The significance of changing gas depends on how that particular meter produces its reading and on the conversion applied by its electronics.
- Range and configuration
- A reading in standard L/min is tied to stated reference conditions. An actual-volume reference can only be compared after the conditions are reconciled.
- Manufacturers and models
- Alicat supplies gas flow instruments in multiple full scales. A gas selection and the reference temperature and pressure belong with the meter identification.
- What we check
- Compare the readings at steady flow, recording the gas and any compensation selected. Use separate points on either side of a range change if the process crosses one.
- Reference equipment and accuracy
- Use a calibrated gas-flow standard suitable for the gas. Temperature, pressure, leaks and gas composition can govern uncertainty at low flow.
- When to repeat the check
- Review drift after a change of gas, contamination or filter work.