Gauge pressure transmitter calibration

A vented gauge-pressure transmitter follows local atmospheric pressure. A sealed-gauge model has a sealed reference and can respond differently as the weather changes. Similar housings can conceal different versions, so include the full model number and suffix.

Send an enquiry
Gauge pressure transmitter — equipment illustration
Equipment illustration
On this page

Lower range value and pressure reference

State the configured range, reference type, output and supply arrangement. If the instrument is ranged above zero, include the actual lower range value alongside the desired output so the comparison follows the setting in service.

A linear 4–20 mA transmitter ranged from 2 to 10 bar should produce its midpoint output at 6 bar. A check calculated on the assumption of a 0-to-10 bar range would point to a false scaling error.

Specifications and service checks

Suppose a vented reference transmitter has a weatherproof enclosure around its cable termination. If readings move after the enclosure is closed, describe the reference vent arrangement as well as the applied pressure. Trapped pressure at the reference can affect the reading.

Range and configuration
Specify the sensor limits and the configured lower and upper values separately. On a linear 4–20 mA range, the mid-span output is 12 mA, even when zero pressure is outside the configured span.
Manufacturers and models
Emerson's Rosemount 3051 family includes differential, gauge and absolute versions with several output options. Use the complete model code.
What we check
Compare the output at applied gauge pressures with its reference side in the specified condition. A blocked atmospheric vent can appear as a changing zero offset.
Reference equipment and accuracy
Use a calibrated pressure reference of the correct mode. A calibrated current meter is also needed for a loop-output check. Both uncertainties enter the result.
When to repeat the check
Review the interval against drift in the working range. Check again after a sensor change, rerange or repair when it affects the reported value.

Technical references Product Data Sheet: Rosemount 3051 Pressure Transmitter