On this page
Meter readings and control lots
For a control check, send the engineering record with the level, lot and assigned interval alongside the value shown on the meter. If two wards have identical models, use each meter’s own identifier. Keep the original display unit on the result. That allows the biomedical team to compare like readings after a change of strips or control material without using patient records.
Glucometer quality-control enquiries
A control-material result, an analytical comparison and a meter’s internal check provide different evidence. Request review of the applicable method and manufacturer requirements. Do not send patient readings or assume that a general electrical calibration establishes blood glucose measurement performance.
Specifications and service checks
Two meters can display different units even when they are from the same maker. Copy the displayed unit and the control lot's assigned range onto the engineering worksheet before comparing readings from different wards.
- Range and configuration
- The US Guide manual gives a measuring range of 20 to 600 mg/dL. Check the unit and regional model printed on the meter used by the facility.
- Manufacturers and models
- Roche’s Accu-Chek Guide is one example of a meter and strip system.
- What we check
- Review control results with the matching strips and control material. A comparison against a laboratory method requires the facility’s approved comparison protocol.
- Reference equipment and accuracy
- Use specified controls within their stated ranges and shelf life. A control check has a different purpose from establishing metrological traceability for a laboratory reference result.
- When to repeat the check
- Follow the meter and strip instructions for checks on a new lot or after storage problems. Use the facility’s quality programme for periodic comparisons.
Technical references Accu-Chek product reference · Accu-Chek technical manual