High speed centrifuge calibration

A high speed centrifuge may have several selectable rotor profiles. The speed displayed in revolutions per minute and a calculated centrifugal force rely on different information. The latter also depends on the rotor geometry selected in the controller.

Send an enquiry
High speed centrifuge — equipment illustration
Equipment illustration
On this page

Drive speed with the fitted rotor

Identify the fitted rotor and the speed points used in the equipment procedure. For models with restricted access to rotational measurement, provide the approved service interface details in advance. A report should name the reference measurement and the display mode it was compared with.

A legacy equipment sheet may show an entry in g while a newer service report quotes revolutions per minute. A direct numerical comparison would be meaningless. Identify the rotor and radius convention behind the acceleration entry before using the two records to assess a change.

Specifications and service checks

A maintenance log showing the time to reach speed can change even when steady rpm stays within its usual band. Separate the acceleration time from the speed held later in the run. Include the model of the rotor used for the observation.

Range and configuration
The drive and the fitted rotor have separate speed limits. Indicated rpm and relative centrifugal force are different quantities. Refrigerated models also have a separate temperature channel.
Manufacturers and models
Eppendorf’s 5804 and 5804 R are benchtop examples. Beckman Coulter’s Avanti family includes floor-standing systems with rotor-specific limits.
What we check
Compare indicated rotational speed with a traceable speed reference using the manufacturer’s approved service access. Identify the fitted rotor and test any timer or temperature channel separately.
Reference equipment and accuracy
Use a speed reference appropriate to the manufacturer’s approved access and speed rating. A control display showing derived relative force also depends on the radius configured for the fitted rotor.
When to repeat the check
Review after a drive or speed sensor repair. Rotor condition and lifetime inspections belong in the maintenance record alongside speed calibration.