High voltage divider calibration

A high-voltage divider provides a reduced output related to the applied high voltage by its ratio. Its output lead and measuring impedance may be part of the calibrated configuration, particularly for AC or rapidly varying signals.

Send an enquiry
High voltage divider — equipment illustration
Equipment illustration
On this page

Divider ratio and connected output instrument

Give the divider identifier, rated voltage, assigned ratio, output load and frequency range or DC polarity. Specify which cable and readout will be used. A ratio reported at one condition should be identified as such when the divider is fitted to another system.

A divider and its readout may each carry an asset number. If the laboratory reports the ratio of the divider, the user applies that ratio to the readout. A certificate for an already scaled display instead gives the end-to-end indication. Mark which result your calibration register is meant to hold.

High voltage dividers and their matched readouts

Identify the divider ratio, rated input, AC or DC use and the readout or load connected to its low-voltage output. Include cable identity and previous assigned ratio values. Ratio and voltage-dependence measurements are made with the divider connected to the specified readout and load.

Specifications and service checks

For a divider nominally marked 1000:1, an input of 1 kV corresponds nominally to 1 V at its output. The readout impedance and cable configuration matter to the working ratio. Identify the components used as part of the measuring chain.

Range and configuration
A 1,000:1 divider produces a nominal 1 V per 1 kV at its design load. Its AC response and permissible voltage depend on the individual rating.
Manufacturers and models
Ross Engineering supplies high-voltage measuring systems and matched probe/meter pairs. Record the divider ratio and the readout input specified for the assembly.
What we check
Determine the ratio with the specified readout impedance and at the required voltage levels. For AC or transient work, establish the frequency response relevant to that use.
Reference equipment and accuracy
Use a calibrated high-voltage reference divider and suitable measurement system. Loading, leakage and frequency response affect the ratio uncertainty.
When to repeat the check
Recheck the pair after readout replacement, damage to insulation or repair to the divider.