High voltage probe calibration

A meter connected to a high-voltage probe reads only a fraction of the voltage at the probe tip. The marked ratio assumes a particular input impedance. For a scope probe, compensation can also affect what the trace shows as frequency changes.

Send an enquiry
High voltage probe — equipment illustration
Equipment illustration
On this page

The probe ratio and its paired meter

A marking of 1000:1 gives the nominal ratio. For example, a report can show the measured ratio of the probe at a DC test point, or the voltage indicated by the particular probe and meter together. Which result does the engineer use? Send the probe model, attenuation setting, readout and the accessories specified for that probe. If the job covers AC, tell us the working frequencies as well.

High voltage probes

Give the attenuation setting, connector, voltage type and requested frequency if applicable. A probe alone and the probe-plus-meter combination are separate report options. RTS calibrates the probe at its specified loading and voltage conditions. Identify this as a probe enquiry rather than assuming the meter's own calibration covers its external attachment.

Specifications and service checks

An attenuating probe may be designed for a particular input impedance at the readout. A replacement display with a different input can change the assembled ratio. Give the intended readout model and interface before choosing a complete-system comparison.

Range and configuration
The nominal ratio means 10 kV would give 10 V at the readout under the probe's stipulated conditions. Identify the meter used with the probe.
Manufacturers and models
Fluke 80K-40 is a 1,000:1 probe designed for use with the specified input impedance. Its voltage and application ratings apply to the complete assembly.
What we check
Compare the probe/readout indication with a traceable high-voltage reference at representative working points with the prescribed load.
Reference equipment and accuracy
Use a calibrated reference divider and appropriate readout. Input resistance, insulation condition and voltage dependence all influence the result.
When to repeat the check
Check following insulation damage, connector replacement or any change in the paired meter specification.

Technical references Fluke 80K-40 High Voltage Probe